Lesson 2 · 30 min
The glass pH electrode and ion-selective electrodes
Why does a thin glass membrane respond to H⁺? Two-point calibration, slope efficiency, sodium and acid errors — and the LaF₃ electrode with TISAB that measures fluoride in drinking water.
In this lesson
- Explain why a glass electrode gives a Nernstian response to H⁺ and how a combination electrode is built
- Calibrate a pH meter with two buffers, compute the slope efficiency and judge when to recalibrate or replace the electrode
- Use the Nikolsky–Eisenman equation to explain the sodium error and the selectivity coefficients of ion-selective electrodes
- Design a fluoride-electrode measurement: why TISAB is added and at what pH to measure
Dip a pH meter into a mixture of 0.10 M NaOH and 0.90 M NaCl and the display says 11.0 — yet the solution’s pH is 13.0. The meter is not broken and the buffers have not expired; the glass membrane has been fooled by a flood of . The glass electrode is the most widely used chemical sensor in the world, with millions of measurements a day relying on it. Knowing how it works and when it goes wrong underpins every pH measurement.
Predict first
A glass electrode’s real slope is only 95 % of the theoretical value. You “calibrate” only the zero with the pH 7.00 buffer, and the meter assumes a 100 % slope. Measuring a sample of true pH 4.00, what does it show?
Pick an answer and lock it in before you look. A wrong guess teaches more than no guess.
How the glass membrane senses H⁺
The sensing part of a pH electrode is a thin glass bulb about 0.1 mm thick. Glass is a network of silicate tetrahedra whose negative charges are balanced by cations such as and . The inner and outer surfaces carry hydrated gel layers tens of nanometres thick, where sites exchange with from the solution: the more acidic the outside, the more binds there, and a potential difference builds across the membrane. Inside the glass, carries the current; itself never crosses.
The bulb holds a chloride buffer of fixed pH and an Ag/AgCl internal reference wire; a combination electrode also packs the external reference (another Ag/AgCl with KCl filling) into the same body. The electrode’s potential is a straight line in sample pH with a slope close to the Nernst slope and an intercept (the zero point, or asymmetry potential) that varies between electrodes and drifts with use — which is why it must be calibrated with buffers every time.
- zero point: electrode potential at pH 7
- electromotive efficiency (slope per cent)
- Nernst slope, 59.16 mV at 25 °C
Calibrate properly: use two (or three) buffers that bracket the sample — 7.00 and 4.01 for acidic samples, 7.00 and 10.01 for basic ones; rinse with distilled water and blot between solutions; wait for a stable reading; keep samples and buffers at the same temperature or use automatic temperature compensation. With one buffer only, the meter can only assume a 100 % slope, and the error grows with distance from the calibration point.
Selectivity: the Nikolsky–Eisenman equation
No electrode responds to one ion alone. When is tiny and large, takes over the exchange sites, the electrode mistakes it for and reads low — the alkaline (sodium) error. In very strong acid the membrane’s response saturates and it reads high — the acid error.
The Nikolsky–Eisenman equation converts each interfering ion j into an “equivalent” amount of the measured ion: the smaller , the more selective the electrode. General-purpose glass has near 10⁻¹¹, which gives an error of about −1 unit at pH 12 in 1 M ; lithium-based “low-sodium-error” glass cuts it below 0.1.
- activity and charge of the measured ion
- activity and charge of an interfering ion
- selectivity coefficient; smaller is better
Ion-selective electrodes swap the glass for a membrane sensitive to another ion: a single crystal of doped with for ; a pressed pellet for and ; a PVC membrane holding valinomycin for (selectivity against about 10⁻⁴); a gas-permeable membrane over a pH electrode for and . The , , and results of a clinical blood-gas analyser are measured this way.
The fluoride electrode suffers almost only from (), but the sample pH must be right: below pH 5 fluoride is protonated to HF (p 3.17) and the electrode cannot see it; above pH 8 hydroxide impersonates fluoride. So samples are mixed 1 : 1 with total ionic strength adjustment buffer (TISAB): concentrated NaCl fixes the ionic strength (so standards and samples share activity coefficients), an acetate buffer holds pH 5–5.5, and CDTA frees fluoride bound to and .
Worked example
Fluoride in drinking water with an ISE
Three fluoride standards (mixed 1 : 1 with TISAB like the sample) read +127.0 mV at M, +68.2 mV at M and +9.4 mV at M. A water sample reads +80.5 mV. Find its fluoride concentration in mg/L.
Concept check
A sample is expected to be near pH 9.3. Which two buffers should you calibrate with?
Try the calculation
Two-point calibration: the pH 7.00 buffer reads −3.5 mV and the pH 4.01 buffer +168.4 mV (25 °C). What is the electrode’s slope efficiency in per cent?
Key takeaways
- A glass electrode’s potential is linear in pH with a slope near 59.16 mV (25 °C); its zero and slope vary, so it must be calibrated with buffers.
- Two-point calibrations must bracket the sample; a slope efficiency of 95–102 % and a zero within ±30 mV are normal.
- Nikolsky–Eisenman: interferents count as ; at high pH and high the alkaline error appears.
- Fluoride electrodes are used in TISAB at pH 5–5.5, avoiding HF formation and interference and fixing the ionic strength.
Data and references
- D. C. Harris, Quantitative Chemical Analysis, 9th ed. (2016), Ch. 15 (the glass electrode, errors in pH measurement, ion-selective electrodes).
- R. P. Buck et al., “Measurement of pH. Definition, standards, and procedures (IUPAC Recommendations 2002)”, Pure Appl. Chem. 74, 2169–2200 (2002).
- R. G. Bates, Determination of pH: Theory and Practice, 2nd ed. (Wiley, 1973).
- Y. Umezawa, P. Bühlmann, K. Umezawa, K. Tohda, S. Amemiya, “Potentiometric selectivity coefficients of ion-selective electrodes”, Pure Appl. Chem. 72, 1851–2082 (2000).
- M. S. Frant, J. W. Ross, “Electrode for sensing fluoride ion activity in solution”, Science 154, 1553–1555 (1966).
- WHO, Guidelines for Drinking-water Quality, 4th ed. (2017): fluoride 1.5 mg/L; GB 5749-2022, Standards for drinking water quality: fluoride 1.0 mg/L.