Lesson 5 · 30 min
Calibration curves and detection limits
Instruments give signals; concentrations come from a calibration line. Least-squares fitting, the unknown’s confidence interval, residual plots, detection and quantitation limits — and standard addition to beat matrix effects.
In this lesson
- Fit a calibration line by least squares and give the uncertainties of its slope and intercept
- Find an unknown concentration and its confidence interval from the calibration line
- Check linearity with a residual plot and calculate the detection (3s/m) and quantitation (10s/m) limits
- Recognise matrix effects and get the right answer by standard addition
A water report says “lead: not detected”. That does not mean there is not a single lead atom; it means the concentration is below the smallest amount the method can reliably tell apart from nothing — its detection limit. Spectrophotometers, atomic-absorption instruments and chromatographs do not report concentrations; they report signals, and a calibration curve built from standards of known concentration turns signal into concentration. This lesson covers how to fit that line, how to give an unknown’s concentration with its uncertainty, and how small a signal can be before it stops meaning anything.
Predict first
A calibration uses six standards from 0 to 5 mg/L. Where must the unknown fall for its concentration, read from the line, to be most precise?
Pick an answer and lock it in before you look. A wrong guess teaches more than no guess.
The least-squares line
Least squares chooses the line that minimises the sum of the squared vertical deviations (the residuals). It assumes that the concentrations x are exact, that all the error is in the signals y, and that it is the same size for every point. Besides the slope and intercept it gives three uncertainties:
- standard deviation of the points about the line
- standard deviations of the slope and intercept
- number of calibration points
- scatter of the unknown’s own readings: shrinks with more readings
- uncertainty in where the whole line sits
- grows the further you are from the centre of the calibration
Worked example
A four-point calibration
Four standards: (1, 2), (3, 3), (4, 4), (6, 5). Find the least-squares line and its uncertainties; an unknown read once gives y = 3.50 — find its x with a 95 % confidence interval.
Residuals, detection and quantitation limits
After fitting, always look at the residual plot: the residuals should scatter randomly about zero. An arc (negative at both ends and positive in the middle, or the reverse) means the response has curved — stray light in a spectrophotometer reading low at high absorbance, say — and you should shorten the linear range rather than trust a handsome R².
When a signal is close to the blank, ask whether it can be told apart from the blank at all. Measure the blank (or a very low sample) many times and find its standard deviation s:
- detection limit LOD = 3s/m: a signal 3s above the blank has only about a 1 % chance of being a false “detected”;
- quantitation limit LOQ = 10s/m: above it the relative standard deviation is about 10 %, good enough to report a number.
Results between the two can only be reported as “detected, below the quantitation limit”.
Matrix effects and standard addition
Everything in a sample except the analyte — salts, organic matter, acidity, viscosity — is the matrix. The matrix can suppress or enhance the signal, so the same concentration gives a different slope in the sample than in pure solvent, and an external calibration with standards in pure solvent comes out systematically low or high. Standard addition spikes known amounts of standard into the sample itself, so the calibration happens in the sample’s own matrix: plot the signal against the amount added, extrapolate the line to zero signal, and the magnitude of the x-intercept, , is the unknown’s concentration.
Concept check
The matrix of a river-water sample reduces the analyte’s signal to 70 % of that in pure solvent. What happens if you calibrate with standards in pure solvent?
Try the calculation
Ten readings of a reagent blank give an absorbance standard deviation of 0.0021; the calibration slope is 0.1988 L/mg. What is the detection limit in mg/L?
Key takeaways
- The least-squares line gives m ± sₘ and b ± s_b; residuals should be random, and a curve means the linear range has been exceeded.
- The unknown’s uncertainty sₓ₀ is smallest at the centre of the calibration and shrinks with more readings (k); the interval is x₀ ± t·sₓ₀ (ν = n − 2).
- Detection limit LOD = 3s/m and quantitation limit LOQ = 10s/m, with s the blank’s standard deviation.
- When the matrix changes the sensitivity, external calibration is biased; standard addition calibrates in the sample’s own matrix and reads the concentration from the x-intercept.