Two-point calibration of a pH meter and sample measurement
A glass combination electrode with an unknown zero and slope: calibrate it with pH 4.01, 7.00 and 10.01 buffers, measure acetic acid, an ammonia buffer, a high-sodium caustic solution and an unknown buffer, then reveal the electrode to find which readings were pulled off by extrapolation or by the sodium error.
Aim Calibrate a pH meter correctly, work out the electrode’s slope efficiency and zero, and measure the pH of an unknown buffer.
Procedure
- Put the electrode in the pH 7.00 buffer, wait for a stable reading and press “Calibrate on this buffer”; note the zero (mV).
- With this one point only, measure samples A (acetic acid) and B (ammonia buffer) and record the displayed pH.
- Complete a two-point calibration with the pH 4.01 buffer and read the slope efficiency; measure A and the unknown again. Recalibrate with 7.00 and 10.01 before measuring the basic samples B and C.
- Switch to the mV display and watch how the reading settles when the electrode moves between solutions.
- Reveal the true pH values and the electrode parameters, compute the error of each reading and explain where it came from.
What to notice
- With one buffer only, the error grows with distance from pH 7: the meter assumed a 100 % slope.
- After a two-point calibration, samples lying between the two buffers read within 0.01 of the truth.
- 0.10 M NaOH + 0.90 M NaCl has a true pH of 13.0, yet the general-purpose glass shows only about 11 — the sodium error, which no calibration can remove.
- Slope efficiencies lie between 95 and 100 % and zeros between a few and a dozen or so millivolts — every electrode differs.
In a real lab
- Rinse the electrode with distilled water and blot it gently between solutions; never wipe the bulb, which charges and scratches it.
- Measure buffers and samples at the same temperature or use automatic temperature compensation; opened buffers absorb CO₂ — the alkaline one especially — and must be replaced regularly.
- Store the electrode in 3 M KCl storage solution, not distilled water; a clogged junction makes readings drift and respond slowly.
- Use a low-sodium-error electrode for strongly alkaline, high-sodium samples, and clean the membrane after protein- or oil-rich samples.