Measure the half-life of barium-137m with a Geiger counter
Elute a few drops containing ¹³⁷ᵐBa from a ¹³⁷Cs/¹³⁷ᵐBa isotope generator, count them with a Geiger counter interval after interval, subtract the background, plot ln(count rate) against time and get the half-life — with its uncertainty — from the slope.
Aim Determine the half-life of ¹³⁷ᵐBa and estimate the uncertainty of the result from counting statistics.
Procedure
- Before any sample is present, count the background for 300 s and find the background rate and its uncertainty √N/t.
- Elute the generator (t = 0) and put the dish of eluate in a fixed position under the counter window.
- Choose a counting interval (say 20 s) and count interval after interval, recording at least 10 of them (about four half-lives).
- For each interval subtract the background, work out the net count rate and its uncertainty, and plot ln(net rate) against the midpoint of the interval.
- Fit a weighted straight line to get the slope −λ, then t½ = ln 2/λ; finally reveal the accepted value and see how many standard uncertainties you are off.
What to notice
- Every count fluctuates: a count N has a standard deviation of about √N. Early counts are large with small relative errors; late counts are small and their error bars grow.
- ln(net rate) against time is a straight line — the signature of first-order decay; the slope does not depend on the starting count rate.
- The last intervals are nearly all background, and after subtraction may even go negative and must be dropped: longer counting helps, but not once the source has faded into the background.
- Repeat with a new sample: the results agree within their uncertainties. Changing the interval from 10 s to 30 s hardly changes the fitted uncertainty.
In a real lab
- The ¹³⁷Cs/¹³⁷ᵐBa generator is a sealed teaching source; the eluate holds ¹³⁷ᵐBa and a trace of ¹³⁷Cs. Wear gloves, work over a tray, and collect the waste according to your school’s rules for radioactive sources.
- At high count rates the Geiger tube’s dead time (about 100 µs) loses counts and makes early rates too low; increase the distance to the tube or apply a dead-time correction.
- Keep the sample in exactly the same position throughout: any change in geometry changes the count rate for reasons that have nothing to do with decay.
- ¹³⁷ᵐBa emits 662 keV γ rays, which a Geiger tube detects with an efficiency of only a few per cent, so the count rate is far below the sample’s actual activity.