Lesson 1 · 25 min

Volumetric analysis: primary standards and standardisation

The 0.1 M on the label is never 0.1000 M. Standardise titrants against primary standards such as KHP, master burettes and pipettes, and see where the uncertainty comes from.

In this lesson

  • Distinguish primary from secondary standards and list what makes a good primary standard
  • Standardise a titrant against a primary standard and compute its concentration from mass and volume
  • Use burettes, pipettes and volumetric flasks correctly, knowing class A tolerances and how to read them
  • Plan direct, back and displacement titrations and estimate the uncertainty of the result

The bottle says “0.1 mol/L NaOH”, but it is never 0.1000 mol/L. Solid NaOH picks up water and CO₂ from the air, and the solution drifts over a few days. Titrations reach an accuracy of one part in a thousand not because of labels but because of a strict routine: a substance of very high purity and exactly known composition — a primary standard — is used to standardise the titrant, and the standardised titrant then measures the unknowns.

Predict first

Titrating with about 0.1 M NaOH, you want to use about 40 mL. Which of these primary standards gives the smallest relative weighing error? (The analytical balance reads ±0.1 mg.)

Pick an answer and lock it in before you look. A wrong guess teaches more than no guess.

Primary and secondary standards

A good primary standard is at least 99.9 % pure with a certificate; matches its formula exactly (water of crystallisation included); is stable in air, not hygroscopic and easy to dry; has a large molar mass; reacts quickly with a definite stoichiometry; and is affordable. The usual ones:

  • for bases, potassium hydrogen phthalate (KHP); for acids, anhydrous sodium carbonate and borax;
  • for redox, potassium dichromate (weighed straight into a standard solution), sodium oxalate (for KMnO₄), potassium iodate (for Na₂S₂O₃);
  • for EDTA, calcium carbonate or zinc metal; for AgNO₃, sodium chloride.

NaOH, HCl, KMnO₄ and Na₂S₂O₃ cannot be weighed into standard solutions; they are made up roughly and then standardised — they are secondary standards.

cT=mS/MS⋅νVTc_\mathrm{T} = \frac{m_\mathrm{S}/M_\mathrm{S}\cdot\nu}{V_\mathrm{T}}
KHP reacts 1 : 1 with NaOH (ν = 1); Na₂CO₃ titrated to its second equivalence point takes 2 H⁺ (ν = 2); sodium oxalate and KMnO₄ react 5 : 2 (ν = 0.4).
cTc_\mathrm{T}
titrant concentration, mol/L
mS,MSm_\mathrm{S}, M_\mathrm{S}
mass and molar mass of the standard
ν\nu
moles of titrant per mole of standard
VTV_\mathrm{T}
volume of titrant at the endpoint, L
Virtual experiment · exact charge-balance solver

Acid–base titration bench

25.00 mL KHCX8HX4OX4\ce{KHC8H4O4} 0.1000 M ← NaOH\ce{NaOH} 0.1000 M

05101520253035404550STIR501001502004.19pH25.0 °CCALREADMODE
Burette reading 0.00mL
pH 4.19
Colourcolourless
0 recorded

Titration curve: the points you recorded

Equivalence 1 0510152025303540Titrant volume V / mL02468101214pH

The endpoint is where the colour change persists. Switch to single drops as you approach it.

pH is re-solved from the exact charge balance after every drop (ideal solution, 25 °C). The flask colour comes from each indicator form’s absorption band via Beer–Lambert and CIE colorimetry. The flash of colour that appears and fades near the endpoint is the local, briefly higher pH where the drop lands, diluted as the stirrer mixes it. Particle proportions are quantitative; their positions are illustrative.
KHP is a weak acid (pKa₂ = 5.41) with an equivalence pH near 9, hence phenolphthalein. Switch to “Na₂CO₃ with HCl”: there are two breaks, and HCl is standardised at the second (methyl red or bromocresol green), boiling off CO₂ near the endpoint to sharpen it.

Volumetric glassware and technique

Class A tolerances at 20 °C: 50 mL burette ±0.05 mL, 25 mL pipette ±0.03 mL, 250 mL volumetric flask ±0.12 mL. Key points:

  • Rinse burettes and pipettes three times with small portions of the solution they will hold, and clear the air bubble from the tip; the titration flask needs only distilled water and may stay wet — the amount of analyte inside does not change.
  • Read the bottom of the meniscus at eye level to 0.01 mL; a drop is about 0.05 mL, and near the endpoint you can add half drops (let a drop hang on the tip and rinse it in).
  • Start near 0.00, read both the initial and final levels; their difference is the volume.
  • The endpoint differs from the equivalence point by the indicator error; subtract a blank titration when needed.

Besides direct titration there is back titration — add a known excess of reagent and titrate what is left, useful when the reaction is slow, the endpoint poor or the sample insoluble (calcium carbonate, the Volhard method); displacement titration — the analyte releases an equivalent amount of another species, which is titrated (calcium via Mg–EDTA); and indirect titration — titrating a product quantitatively related to the analyte (iodometry).

Worked example

Standardise NaOH, then a back titration of an antacid

(a) 0.8154 g of KHP (M = 204.22 g/mol) needs 39.62 mL of NaOH; find c(NaOH). (b) 0.2500 g of antacid powder reacts with 50.00 mL of 0.1000 M HCl, and the excess acid takes 12.40 mL of 0.1000 M NaOH. Find the mass fraction of CaCO₃.

Concept check

You fill a burette that still has distilled water on its walls (not rinsed with titrant) with standard NaOH and titrate an acid. The acid concentration you calculate will be:

Try the calculation

0.4912 g of anhydrous Na₂CO₃ (M = 105.99 g/mol) is titrated to its second equivalence point with 45.26 mL of HCl. What is c(HCl)?

mol/L

Key takeaways

  • Primary standards are pure, of exact composition, stable, non-hygroscopic and of large molar mass; NaOH, HCl, KMnO₄ and Na₂S₂O₃ are secondary standards that must be standardised.
  • c_T = (m/M)·ν/V: choose a mass that uses most of the burette so that both weighing and reading errors stay small.
  • Rinse burettes and pipettes with the solution they will hold; the flask only with water; read to 0.01 mL.
  • Besides direct titration there are back, displacement and indirect titrations.
Progress is stored only in this browser and never uploaded.

Data and references

  1. D. C. Harris, Quantitative Chemical Analysis, 9th ed., W. H. Freeman (2016), ch. 2 (glassware, calibration) and ch. 7, 10–11 (titrations, primary standards).
  2. D. A. Skoog, D. M. West, F. J. Holler, S. R. Crouch, Fundamentals of Analytical Chemistry, 9th ed. (2014), ch. 13–14.
  3. ASTM E287/E288/E969 class A tolerances, as tabulated in Harris, Tables 2-2 to 2-4.