Sorting metals with a hand-held XRF analyser
Excite X-ray fluorescence from stainless steel, brass, solder, a gold ring and a handful of contaminated soil, assign every peak to an element from NIST line energies — and meet the notorious overlap of lead and arsenic.
Aim Identify every element in each sample — no more and no fewer — from the positions of its Kα, Kβ, Lα and Lβ lines.
Procedure
- Choose “stainless steel”. Start with the element you are surest of and see which peaks its lines land on.
- Keep adding elements until every peak is accounted for and nothing is left over.
- Analyse the brass, the solder and the gold ring in turn. Notice that heavy elements show only L lines: their K lines lie beyond what a 40 kV tube can excite.
- Finally analyse the contaminated soil. Choose lead alone, then arsenic alone, and see which peak each leaves unexplained.
What to notice
- The stainless steel has six peaks: chromium at 5.41 and 5.95 keV, iron at 6.40 and 7.06 keV, nickel at 7.48 and 8.26 keV — a Kα–Kβ pair for each element.
- The tin in the solder shows K lines at 25.3 and 28.5 keV as well as L lines near 3.4 keV; the lead shows only its L lines at 10.55 and 12.61 keV.
- In the soil the peak at 10.55 keV could be lead Lα₁ or arsenic Kα₁: only the peaks at 12.61 keV (lead Lβ₁) and 11.73 keV (arsenic Kβ₁) show that both are present.
- Add manganese by mistake: its Kα (5.90 keV) lands next to chromium Kβ (5.95 keV) but 48 eV away, and its Kβ has no peak — the analyser reports that there is no manganese.
In a real lab
- A hand-held XRF analyser emits X-rays: keep the window pressed against the sample, never point it at anyone, and follow the instrument’s interlocks and dose rules.
- XRF sees only the surface layer: plating, paint and scale make the reading describe the surface rather than the metal, so clean the spot first.
- In air the low-energy lines of light elements such as magnesium, aluminium and silicon are lost; quantitative results need calibration with certified alloys.