Dissolved oxygen by the Winkler method
Fix the oxygen in a water sample as a brown precipitate, release iodine with acid and titrate it with thiosulfate. Compare cold, warm and undersaturated samples and see how dissolved oxygen depends on temperature.
Aim Measure the dissolved oxygen of air-saturated water by the Winkler method, compare it with the Benson–Krause saturation value, and work out the percentage saturation from the titre.
Procedure
- Sample: 20 °C, air-saturated. Press “Add MnSO₄ and alkaline iodide–azide” and then “Add conc. H₂SO₄”, watching the colour of the precipitate and of the solution.
- Titrate 201.3 mL (equal to 200 mL of original sample) with 0.02500 M thiosulfate. When the solution has faded to pale yellow, add starch and continue drop by drop until the blue just disappears; record the endpoint reading.
- Repeat with air-saturated samples at 5 °C and at 30 °C.
- Finally measure a sample at 20 °C that is only 60 % saturated, and one with all its oxygen removed (0 %).
What to notice
- Air-saturated water at 20 °C: a brown precipitate, yellow-brown after acidification, and an endpoint near 9.09 mL — 9.09 mg/L, matching the Benson–Krause saturation value.
- The 5 °C sample needs about 12.77 mL and the 30 °C sample only about 7.56 mL: between 5 and 30 °C the oxygen water can hold falls by about 41 %.
- The 60 % sample needs about 5.46 mL (5.46 mg/L); the deoxygenated sample gives white Mn(OH)₂ on fixing, is almost colourless after acidification, and does not turn blue with starch.
In a real lab
- When sampling, lead the water to the bottom of the bottle through a tube and let it overflow for a while before stoppering, with no bubble left inside; fix the oxygen on the spot and titrate soon after acidifying, before iodine evaporates or air oxidises the iodide.
- The alkaline iodide–azide reagent is strongly caustic and sodium azide is highly toxic; add the concentrated sulfuric acid slowly down the neck of the bottle. Wear gloves and eye protection.
- Thiosulfate solutions slowly decompose and must be standardised against potassium iodate or dichromate before use; old starch solution loses its sensitivity and blurs the endpoint.