Manganese in steel by spectrophotometry
Oxidise the manganese in a steel sample to purple permanganate, compare the spectra of a standard and the steel solution to choose the analytical wavelength, then calibrate with KMnO₄ standards and report the steel’s manganese content. The steel’s chromium quietly spoils the wrong wavelength.
Aim Determine the mass fraction of manganese in a steel by external calibration, and show how the choice of analytical wavelength avoids interference from chromium (as dichromate).
Procedure
- Note the sample number and mass: about 0.5 g of steel has been dissolved in nitric–phosphoric acid, its manganese oxidised to MnO₄⁻ with potassium periodate, and the solution made up to 250.0 mL.
- Compare the spectra of a KMnO₄ standard and the steel solution and choose the analytical wavelength from 440, 480, 526, 546 and 580 nm.
- Make five standards of 1–5 × 10⁻⁴ M from 1.000 × 10⁻³ M KMnO₄ stock and measure them and two or three reagent blanks at the chosen wavelength.
- Measure the steel solution 3–5 times, read its MnO₄⁻ concentration with a confidence interval from the calibration line, and convert to the manganese mass fraction of the steel.
- Reveal the true value, then repeat at another wavelength (440 nm, say) and compare.
What to notice
- KMnO₄ has two almost equal peaks at 526 and 546 nm; near 440 nm the steel solution absorbs noticeably more than the standard — that extra is Cr₂O₇²⁻.
- At 440 nm the calibration slope is small and the steel result comes out clearly high; at 546 nm chromium hardly absorbs and the result is accurate.
- After a change of wavelength every old reading is void: the calibration and the sample must be measured at the same wavelength under the same conditions.
In a real lab
- The nitric acid, phosphoric acid and potassium periodate are corrosive or oxidising: wear goggles and gloves and heat in a fume hood; phosphoric acid also complexes Fe³⁺ and removes its yellow colour.
- Rinse the cell with the solution to be measured and wipe the optical faces; zero on the reagent blank before each series.
- Permanganate slowly fades in the presence of organic matter or reducing agents: make standards and samples fresh and measure them promptly.
- Cr(VI) is toxic and carcinogenic: collect the waste separately for reduction and disposal.