Spectrophotometry: measure the spectrochemical series

Record the absorption spectra of nine nickel(II) and chromium(III) complexes, find the lowest-energy d–d band, convert its wavelength into the splitting Δo and rank the ligands by it.

Difficulty ●●○ ≈ 40 min

Aim Obtain Δo from the lowest-energy band of d³ and d⁸ ions, arrange the ligands in the spectrochemical series and compare the two metals.
Virtual experiment · UV–vis–NIR 350–1300 nm · 1 cm cuvette

Spectrophotometer: measure the spectrochemical series

Absorbance A @ 575 nm 0.017
4005006007008009001000110012001300Wavelength λ / nm00.10.20.30.40.5Absorbance A
No data yet. Pick a wavelength and press “Read and record”.
Absorbance is computed with Beer–Lambert from each complex’s band model (literature band positions, teaching-fit ε and widths), plus a deterministic ±0.003 “instrument noise” hashed from the wavelength and sample, so a repeated reading gives the same value. For Cr(III) and Ni(II) the lowest-energy band equals Δo exactly; the cuvette colours are computed too.

Procedure

  1. Choose 0.10 M [Ni(HX2O)X6]X2+\ce{[Ni(H2O)6]^2+}. Set the monochromator to 400 nm and press “Read and record”; step the wavelength and read again, or press “Scan every 10 nm” to get the whole spectrum from 350 to 1300 nm.
  2. Find the band at the longest wavelength (lowest energy) and add readings every 2–5 nm around its top to pin down the wavelength of maximum absorption, λmax.
  3. Enter λmax: the instrument converts it to ν̃ = 10⁷/λ in cm⁻¹ and to kJ/mol, and compares it with the literature value. Add the result to the table.
  4. Repeat for [Ni(NHX3)X6]X2+\ce{[Ni(NH3)6]^2+} and [Ni(en)X3]X2+\ce{[Ni(en)3]^2+}, then for the six chromium(III) samples (F⁻, H₂O, oxalate, NH₃, en, CN⁻).
  5. Order the ligands by Δo for nickel and for chromium and compare with the spectrochemical series; then compare Δo for Cr(III) and Ni(II) with the same ligand.
  6. For each chromium(III) complex read the second band as well, use the Tanabe–Sugano relations to get the Racah parameter B from the two bands, divide by the free-ion 917 cm⁻¹ to get β, and rank the nephelauxetic series.

What to notice

  • Nickel shows three bands, the lowest in the near infrared: 1176 nm for the aqua ion, moving to 930 nm and 893 nm with NH₃ and en.
  • Chromium shows two visible bands; the lower-energy one moves from 671 nm for F⁻ all the way to 375 nm for CN⁻, and the solutions go from green through violet to yellow-orange and pale yellow.
  • Both metals give the same ligand order: F⁻ < H₂O ≈ oxalate < NH₃ < en < CN⁻.
  • With the same ligand, Δo for Cr³⁺ is about twice that for Ni²⁺: a more highly charged ion draws the ligands closer and splits the d orbitals more.
  • β falls from 0.98 for F⁻ through H₂O 0.79, NH₃ 0.72, en 0.68 and oxalate 0.67 to 0.53 for CN⁻: the softer the ligand and the more covalent its bond to the metal, the further the d-electron cloud spreads and the less the electrons repel each other.
  • The bands are broad, so the top can be located only to within a few nanometres; the “instrument noise” is deterministic, so repeating a reading gives the same value.

In a real lab

  • Zero the instrument on a blank of the solvent (or of the ligand solution alone) before each measurement, and use matched cuvettes; glass and plastic absorb below about 340 nm, so the ultraviolet needs quartz.
  • d–d transitions are weak (ε only 1–100 L mol⁻¹ cm⁻¹), so 0.01–0.1 M solutions are needed. Most teaching instruments stop at 1100 nm; the 1176 nm band of [Ni(HX2O)X6]X2+\ce{[Ni(H2O)6]^2+} needs a near-infrared detector, otherwise Δo has to be found from the second band with a Tanabe–Sugano diagram.
  • Real bands often carry shoulders (spin–orbit coupling, Jahn–Teller distortion), for example near 15 400 cm⁻¹ for [Ni(HX2O)X6]X2+\ce{[Ni(H2O)6]^2+}. Commercial CrClX3 ⋅ 6 HX2O\ce{CrCl3.6H2O} gives green [CrClX2(HX2O)X4]X+\ce{[CrCl2(H2O)4]+}; for violet [Cr(HX2O)X6]X3+\ce{[Cr(H2O)6]^3+}, dissolve chrome alum instead.
  • Nickel salts are sensitisers and possible carcinogens, and ethylenediamine and concentrated ammonia are corrosive and irritating: wear gloves and work in a fume cupboard. Never acidify cyanido complexes (HCN is released); school practicals usually leave them out.