Powder diffraction: identify an unknown metal
A pinch of metal powder, a beam of copper X-rays and one diffraction pattern. From the peak positions, work out the lattice type and the lattice parameter, then find the metal on a reference card.
Aim Use ratios of to decide whether an unknown cubic metal is simple, body-centred or face-centred, calculate its lattice parameter with Bragg’s law, and identify the metal.
Procedure
- Enter a sample number (1–99). The diffractometer scans 2θ = 10°–120° with Cu Kα₁ radiation (λ = 154.06 pm) and lists every peak its peak search finds.
- Read the ratios. Since , they are ratios of . Find the column that turns into whole numbers after multiplying by 1, 2 or 3.
- Face-centred gives 3, 4, 8, 11, 12… (×3); body-centred gives 2, 4, 6, 8, 10… (×2); simple cubic gives 1, 2, 3, 4, 5, 6, 8… — with 7 missing.
- Choose the lattice you have decided on. The table labels each peak and computes . If every peak gives the same , the indexing is self-consistent.
- Compare your mean with the reference card, choose the metal and check. Repeat with a few more sample numbers.
What to notice
- Simple and body-centred cubic give identical ratios for the first six lines (1:2:3:4:5:6); only the seventh line separates them — simple cubic skips 7, body-centred does not. With only six lines, decide by whether matches a real metal.
- Pick the wrong lattice and the values drift from peak to peak — that drift is the test of an indexing.
- High-angle peaks give the most precise : near θ = 90° a given reading error corresponds to the smallest error in .
- Switch to Mo Kα and every peak moves to lower angle, but the indexing and stay the same — the crystal has not changed, only the ruler.
In a real lab
- Real copper radiation contains the Kα₁/Kα₂ doublet, which splits high-angle peaks; software strips Kα₂ or a monochromator removes it.
- Grind the sample fine (about 1–10 μm) and pack it flat with random orientation; coarse grains or preferred orientation badly distort the relative intensities.
- A sample surface that sits off the goniometer axis shifts every peak; precise lattice parameters use an internal standard such as silicon powder.
- X-rays are ionising radiation: diffractometers have interlocked enclosures that keep the beam off while the door is open — never bypass them.