Blood alcohol by headspace gas chromatography with an internal standard
Separate ethanol from the 1-propanol internal standard on a polyethylene glycol capillary column, build an area-ratio calibration line from five calibrators, measure the ethanol in an unknown blood sample and decide whether it crosses the drink-driving and drunk-driving thresholds.
Aim Determine the ethanol concentration (mg/100 mL) of an unknown blood sample by the internal-standard method, and compare the precision of calibrating with the ethanol area alone and with the area ratio.
Procedure
- Prepare the vials: into each headspace vial put 0.50 mL of calibrator (20–200 mg/100 mL ethanol) or blood and 2.00 mL of 1-propanol internal-standard solution (0.100 g/L), and seal at once.
- Inject each of the five calibrators and check the chromatograms: ethanol and 1-propanol must be fully resolved (and acetone, if present, must be clear of ethanol).
- Record the ethanol and propanol areas of every injection and compare two calibration plots: ethanol area alone, and the area ratio.
- Inject the unknown blood sample at least twice, read the ethanol concentration from the area-ratio line and report the mean.
- Reveal the true concentration and compute the relative error; then repeat with another blood sample.
What to notice
- Peak areas differ by a few per cent between injections — the injection volume varies — while the ethanol-to-propanol ratio barely moves.
- The calibration with ethanol area alone has a visibly lower r²; the area-ratio line is almost perfect.
- Per gram, the FID responds to ethanol only about 76 % as strongly as to propanol, so the area ratio is not the concentration ratio — the response factor must be measured with calibrators.
- Some blood samples contain acetone (in diabetic ketosis, for instance); on the polar column it elutes before ethanol and does not interfere.
In a real lab
- Blood is a biohazard: wear gloves and dispose of waste as clinical waste; keep ethanol calibrators tightly sealed to avoid evaporative loss.
- Headspace equilibration needs a fixed temperature (for example 15 min at 60 °C), identical for every sample and calibrator.
- Forensic labs usually run two columns of different polarity in parallel and validate every batch with control and blank samples.
- Report the measurement uncertainty; when a result is near the 20 or 80 mg/100 mL thresholds, the uncertainty decides the conclusion.